FCS Calculator

Fluorescence Correlation Spectroscopy — diffusion time, concentration, and confocal volume.

Model: Textbook approximation

Results

Structure parameter κ = z₀/w₀5.00
Effective volume V_eff0.7517 fL
Molecules in V_eff, N22.63
G(0) = 1/N (diffusion)0.04418
G(0⁺) with triplet0.04909
Diffusion coefficient D450.0 µm²/s
Mean count rate N·ε679 kHz
SNR of G at the shortest lag92.8

Autocorrelation curve

Model

V_eff = π^(3/2) w₀² z₀, N = c·N_A·V_eff, D = w₀²/(4τ_D)

G(τ) = (1/N)(1 + τ/τ_D)⁻¹(1 + τ/(κ²τ_D))^(−1/2) [1 + T/(1 − T)·e^(−τ/τ_T)]

SNR = ε √(Δτ · T_meas) / √(1 + 1/N)

3D Gaussian detection volume (Hess & Webb 2002) with the triplet term of Widengren, Mets & Rigler (1995); T is the mean fraction of molecules in the triplet state, so G(0⁺) = 1/(N(1 − T)). The SNR is Koppel's (1974) shot-noise limit for the shortest lag channel (εΔτ ≪ 1, ε²Δτ τ_D ≪ 1): it grows with the brightness per molecule and, for N ≫ 1, does not depend on N. Background light, afterpulsing and the deviation of a real confocal volume from a Gaussian (larger at high NA or with a mismatched pinhole) are not included.

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

3D Gaussian detection volume with free diffusion and one triplet state. The SNR is Koppel's shot-noise limit at the shortest lag; background, afterpulsing and non-Gaussian confocal volumes are not included.

  1. Hess S. T., Webb W. W. (2002). Focal volume optics and experimental artifacts in confocal fluorescence correlation spectroscopy. Biophys. J. 83, 2300–2317. Effective volume π^(3/2) w₀² z₀ and the 3D Gaussian G(τ). doi:10.1016/S0006-3495(02)73990-8
  2. Widengren J., Mets Ü., Rigler R. (1995). Fluorescence correlation spectroscopy of triplet states in solution: a theoretical and experimental study. J. Phys. Chem. 99, 13368–13379. Triplet factor 1 + T/(1 − T)·e^(−τ/τ_T). doi:10.1021/j100036a009
  3. Koppel D. E. (1974). Statistical accuracy in fluorescence correlation spectroscopy. Phys. Rev. A 10, 1938–1945. SNR set by the counts per molecule, independent of N. doi:10.1103/PhysRevA.10.1938