Plenoptic Camera Design

Plenoptic (light-field) camera design: lenslets and views, microlens f-number matching, refocusable object range, data size and assembly tolerance (Ng 2005).

Model: Textbook approximation

Spatial samples (lenslets)

288 × 288

one sample per 125 µm lenslet

Views N × N

13 × 13 = 169

whole pixels across a micro-image; 61.87 arcmin per pixel at the lenslet

Microlens vs main-lens f-number

f/4.00 vs f/4.00

micro-image 100 % of the pitch: matched: micro-images tile the sensor

Refocusable object range

809 mm – 1.34 m

image side ±N·N_w·p_µ = ±6.50 mm; conventional ±0.500 mm

4D light field

14.02 M samples

28.04 MB at 16 bit; raw frame 32.00 MB

Lenslet-to-sensor tolerance

±36.0 µm

p_px·f_µ/p_µ; lenslet Airy diameter 5.37 µm vs 9.0 µm pixels

Model

A plenoptic camera (Ng et al. 2005) puts a microlens array where the sensor would be and the sensor one microlens focal length f_µ behind it. Each lenslet is one spatial sample, and the N pixels across its micro-image sample the main-lens aperture: N × N views of the scene. The defaults are close to Ng's prototype (125 µm f/4 lenslets, 9 µm pixels), whose micro-images are 12 pixels across (§6.3).

The micro-image of the aperture has diameter f_µ/N_w, so the microlens f-number f_µ/p_µ must equal the main lens's image-side (working) f-number N_w. A larger N_w leaves unused pixels; a smaller one makes neighbouring micro-images overlap, mixing neighbouring views in a band D − p_µ wide, so N counts only the 2p_µ − D clean pixels. The lenslet-to-sensor spacing must be right within p_px·f_µ/p_µ.

Refocusing sums shifted sub-aperture views, each seen through an aperture N times narrower, so a refocused image is sharp (blur within one lenslet) over the depth of field of an f/(N·N_w) lens: ±N·N_w·p_µ on the image side, mapped to object distances by the thin-lens equation. Ng's prototype refocused within the depth of field of f/22 against the ideal f/45, about half this (§6.3).

Spatial–angular trade-off (matched lenslets)

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

Plenoptic 1.0 with the sensor one microlens focal length behind the array. Refocus range: geometric depth of field of an f/(N·N_w) aperture with the lenslet as circle of confusion, through a thin main lens; Ng's prototype reached about half of it. Overlapping micro-images keep only 2p_µ − D clean pixels.

  1. Ng R., Levoy M., Brédif M., Duval G., Horowitz M., Hanrahan P. (2005). Light field photography with a hand-held plenoptic camera. Stanford Tech. Rep. CTSR 2005-02. Sections 3.1, 3.2 and 4.2. graphics.stanford.edu
  2. Levoy M., Ng R., Adams A., Footer M., Horowitz M. (2006). Light field microscopy. ACM Trans. Graph. 25(3), 924–934. Eqs. 1–5 and Table 1. doi:10.1145/1141911.1141976