1. Home
  2. ›Imaging
  3. ›CCD and CMOS Sensor Design

CCD and CMOS Sensor Design

Pixel SNR, dark charge against cooling, ADC step and dynamic range of a camera sensor, with sCMOS and cooled-CCD presets.

Model: Textbook approximation

Load preset:

Signal

700.0 e⁻

QE × 1000 photons

SNR

26.41

Total temporal noise 26.50 e⁻

Dark Charge

0.0111 e⁻

1.114e-1 e⁻/s at 0 °C

Noise Floor

1.511 e⁻

3.02 DN

Dynamic Range (FW/σ_read)

84.44 dB

14.02 stops, the usual datasheet figure

Dynamic Range at this exposure

84.38 dB

14.01 stops: (min(FW, ADC full scale) − D)/noise floor

ADC Full Scale

32768 e⁻

Quantization noise 0.144 e⁻

Peak SNR √FW

158.1

Shot-noise limit at the largest signal

Dark ≤ Read Noise Below

46.0 °C

Cooling further gains little at this exposure

Dark Current Density at 25 °C

0.758 pA/cm²

q·R/p² for a 6.5 µm pixel

Full-Well Density

592 e⁻/µm²

Full well per pixel area

S = QE·N; D = R₂₅·2^((T − 25 °C)/T_d)·t; σ_q = Δ/√12, Δ in e⁻/DN

SNR = S / √(S + D + σ_read² + σ_q²)

DR = 20·log₁₀((S_max − D)/σ_floor), S_max = min(FW, K·(2^bits − 1)), σ_floor = √(σ_read² + D + σ_q²)

Dark ≤ read noise for T ≤ 25 °C + T_d·log₂(σ_read²/(R₂₅·t))

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

Poisson signal and dark charge, Gaussian read noise and quantization noise Δ/√12 (Δ in e⁻/DN) add in quadrature (EMVA 1288). Dark current follows the doubling law about 25 °C, the local form of the n_i scaling. Fixed-pattern noise and nonlinearity near full well are left out; the presets are representative, not a specific camera.

  1. Janesick J. R. (2007). Photon Transfer: DN → λ. SPIE Press, ch. 5. Read, dark and shot noise in quadrature, full well and dynamic range. doi:10.1117/3.725073
  2. EMVA Standard 1288, Release 3.1 (2016). Standard for Characterization of Image Sensors and Cameras, Eq. (11) and (48). Quantization noise σ_q² = 1/12 DN² and the total SNR. emva.org
  3. Janesick J. R. (2001). Scientific Charge-Coupled Devices. SPIE Press. Dark current temperature dependence and dark shot noise. doi:10.1117/3.374903
Report a problem