Jacquinot Advantage

Field of view, field-stop size and étendue of an FTIR interferometer, and its throughput ratio to a grating spectrometer of the same resolving power.

Model: Textbook approximation

Interferometer

Resolution Δν̃ = 1/L

1.000 cm⁻¹

resolving power R = ν̃L = 1000

Max Solid Angle Ω = 2πΔν̃/ν̃

0.006283 sr

half-angle 44.73 mrad (2.56°)

Field (J-) Stop Diameter

8.951 mm

2f tan α at the collimator focus

FTS Étendue A·Ω

0.03084 cm²·sr

beam area 4.909 cm²

Grating Spectrometer at the Same R and Beam Area

Jacquinot Advantage Ω_FTS/Ω_slit

136.0×

π f/(h tan θ); Jacquinot's 2πf/h = 157.1 assumes 2 tan θ = 1

Slit Width Needed for R

288.7 µm

w = 2f tan θ/R (slit-limited; diffraction limit mNρ must exceed R)

The highest wavenumber in the band sets the field stop: a ray at angle θ sees the path difference x cos θ, so its lines appear at ν̃ cos θ; the field is limited to the cone where that shift stays below the resolution.

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

Geometric optics: unapodized resolution 1/L, circular field stop, grating spectrometer in Littrow limited by its slit width; both instruments share the beam area.

  1. Jacquinot P. (1954). The luminosity of spectrometers with prisms, gratings, or Fabry-Perot etalons. J. Opt. Soc. Am. 44, 761–765. doi:10.1364/JOSA.44.000761
  2. Griffiths P. R., de Haseth J. A. (2007). Fourier Transform Infrared Spectrometry, 2nd ed., ch. 7. Wiley.